DATASCOPE CORP. - MAHWAH CARDIOSAVE RESCUE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-83
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
It was reported that prior to transport, the cardiosave intra-aortic balloon pump (iabp) was getting multiple messages: ¿insufficient pressure for vent test¿.A hissing noise was noted from the back of the rescue near the handle.It was suggested to power the pump on and use the refilling station to fill the reservoir.Customer confirmed the helium indicator on the screen was full.When the leak test was run again, the helium was empty and hissing could be heard again.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.E1.Initial reporter was shortened due to character limitations.The complete name is: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE RESCUE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key19006146
MDR Text Key338945963
Report Number2249723-2024-01325
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108407
UDI-Public10607567108407
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-83
Device Catalogue Number0998-00-0800-83
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Is the Device Single Use? No
Type of Device Usage Reuse