MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0700
Device Problem Connection Problem (2900)
Patient Problem Insufficient Information (4580)
Event Date 03/14/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported " iab pump died when unplugged.The user saw stacked alerts of battery life < 20mins, < 10 mins and < 5 mins".Additionally it was reported that "patient did not sustain injury or expire".To complete/continue therapy, the pump was exchanged for another console prior to traveling to the or.The patient's current condition is reported as "critical" please see associated mdr #3010532612-2024-00225.
 
Manufacturer Narrative
(b)(4).The reported complaint of "pump died when unplugged" is not able to be confirmed.No part or recorder strip was returned to teleflex chelmsford for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.Iabp users and servicers must follow the operating instructions manual for recommendations on usage, charging, maintenance and storage of the batteries.If battery maintenance is not performed per the iabp operating instructions manual, the battery may not provide the expected minimum run time of operating power.
 
Event Description
It was reported " iab pump died when unplugged.The user saw stacked alerts of battery life < 20mins, < 10 mins and < 5 mins".Additionally it was reported that "patient did not sustain injury or expire".To complete/continue therapy, the pump was exchanged for another console prior to traveling to the or.The patient's current condition is reported as "critical".Associated mdr #3010532612-2024-00225.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19006241
MDR Text Key339007065
Report Number3010532612-2024-00224
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
AC3; AC3
-
-