Model Number CF-H260AI
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Event Description
It was reported that the colonovideoscope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
The issue was not confirmed, as the scope was not microbiologically tested by olympus.The investigation is ongoing and follow up with the customer is currently being performed.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist, the device evaluation and the lms final investigation.Despite good faith attempts the user culture results were not shared.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.The reported event was not confirmed as the scope was not microbiologically tested by olympus.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19006327
MDR Text Key339047637
Report Number9610595-2024-06697
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCF-H260AI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Was the Report Sent to FDA? No
Date Manufacturer Received03/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
EVIC LUCERA ELITE VIDEO SYSTEM CENTER.
-
-