During a neurovascular procedure physician used a balloon to dilate the target vessel first.The balloon was then deflated and maintained in the target location.The physician then delivered the subject stent through the balloon catheter to the target lesion.The physician withdrew the balloon catheter to deploy the subject stent and after 2-3mm tip of the subject stent was deployed out, it was found that this segment did not expanded.The three markers on the subject stent were not expanded.The operator continued to withdraw the balloon catheter to deploy the subject stent, but resistance was encountered and the subject stent could not be deployed.The operator tried to retract the subject stent but could not retract the subject stent back into balloon catheter.Finally the operator withdrew the subject stent out together with the balloon catheter.The subject stent and balloon catheter were replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|