Catalog Number M003EZAS40210
Device Problems Premature Activation (1484); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
During a neurovascular procedure physician used a balloon to dilate the target vessel first.The balloon was then deflated and maintained in the target location.The physician then delivered the subject stent through the balloon catheter to the target lesion.The physician withdrew the balloon catheter to deploy the subject stent and after 2-3mm tip of the subject stent was deployed out, it was found that this segment did not expanded.The three markers on the subject stent were not expanded.The operator continued to withdraw the balloon catheter to deploy the subject stent, but resistance was encountered and the subject stent could not be deployed.The operator tried to retract the subject stent but could not retract the subject stent back into balloon catheter.Finally the operator withdrew the subject stent out together with the balloon catheter.The subject stent and balloon catheter were replaced and the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS 4.0MM X 21MM NO TIP - CE
Type of Device
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key19006743
MDR Text Key339333735
Report Number3008881809-2024-00137
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P180031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberM003EZAS40210
Device Lot Number24297792
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GATEWAY BALLOON (STRYKER)
Patient Age60 YR
Patient SexMale
Patient Weight60 KG
Patient RaceAsian
-
-