Model Number 106524US
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Skin Infection (4544)
Event Date 06/27/2023
Event Type  Injury  
Event Description
It was reported that on 27jun2023, a new culture of driveline was noted with serratia marcescens 1+, staph aureus 1+, and pseudomonas aeruginosa which was treated with intravenous (iv) antibiotics.The patient's follow-up culture on (b)(6) 2023 was positive for gram negative rods and serratia marcescens 1+.They continued iv antibiotics.Then, follow-up wound culture on 24feb2024 resulted with no growth.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the pump remained in use until the patient underwent a heart transplant on (b)(6) 2024 (see manufacturer report number 2916596-2024-01766).A correlation between the device and the report of infection could not be conclusively determined.Infection is listed in the instructions for use as a potential adverse event that may be associated with the use of heartmate 3 left ventricular assist system}.Infection has been previously investigated and will continue to be monitored through quality data reviews, which are conducted on production and post product signals to evaluate if products are conforming to product requirements.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key19006983
MDR Text Key338980174
Report Number2916596-2024-01852
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public00813024013297
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/17/2022
Device Model Number106524US
Device Lot Number8022548
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age45 YR
Patient SexMale
Patient Weight90 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-