• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PENTA 3MM LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PENTA 3MM LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3228
Device Problem High impedance (1291)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 03/11/2024
Event Type  Injury  
Manufacturer Narrative
Section b3: event date is estimated.Section d4: device serial and lit number are unavailable.Section h4: device manufacturer date is unknown.
 
Event Description
It was reported that the patient was getting ineffective left side coverage from their scs lead.An impedance check indicated that the patient had high impedance on 9 out of the 16 contacts.Surgical intervention may take place at a later date to address the issue.
 
Event Description
Additional information was received indicating that the patient underwent surgical intervention on (b)(6) 2024 wherein the patient's lead was explanted, replaced, and therapy was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key19008585
MDR Text Key338978956
Report Number3006705815-2024-02392
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017253
UDI-Public05415067017253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2024
Device Model Number3228
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS IPG (X1).; SCS LEAD EXTENSION (X2).
Patient Outcome(s) Other;
Patient SexMale
Patient Weight79 KG
-
-