PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation was informed that during the aquablation treatment, the aquabeam robotic system encountered an "e22 - motorpack error" during jet alignment.The procedure setup was executed without any issues, and the system was primed successfully.However, upon attempting to align the jet inside the patient by stepping on the pedal, an e-22 error occurred.Despite efforts to clear the error, it was necessary to undock the handpiece and replace it before proceeding.Subsequently, the procedure was successfully completed after the handpiece replacement.The incident resulted in a surgical delay over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam handpiece was returned for investigation.Investigation into this issue has confirmed the occurrence of "e22 - motorpack" error upon docking and could be cleared.The handpiece was deconstructed and viewed under magnification.No signs of fluid ingress were observed on the sensor board and the encoder wheel.The root cause is undeterminable as the e22 could be cleared.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c01418 was conducted, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were segregated and reworked to address the non-conformance.The handpiece passed final inspection prior to release for distribution.The current user manual sj-um0101-00 rev.B, um, aquabeam robotic system user manual, us, baytech was reviewed and states: table 5: system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
Correction: ab2000-b / serial number 19c01610 and aquabeam handpiece / lot number 23c01418.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c01418 was conducted, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were segregated and reworked to address the non-conformance.The handpiece passed final inspection prior to release for distribution.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose 95134
MDR Report Key19008761
MDR Text Key339245819
Report Number3012977056-2024-00094
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20191205J
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/21/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-