It was reported that boston scientific technical services was contacted for a discussion regarding the elevated, out-of-range pacing impedance measurements (>3000 ohms) and loss of capture from this right atrial (ra) lead.During the data review, out-of-range right ventricular (rv) lead pacing impedance measurements (>3000 ohms) were also observed.As the patient is not pacemaker dependent, the physician elected to reprogram the device sensing mode to resolve the event and no adverse patient effects were reported.Both the ra and rv leads remain implanted and in-service.
|