Model Number 101967-950
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
Senseonics was made aware of an incident where the user received an early sensor replacement alert resulting in early sensor removal.The user received the alert on (b)(6) 2023, on day 48 of sensor life.No adverse events were associated with this complaint.
 
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.A review of the transmitter alert history in data management system (dms) showed that a led disconnect error was first triggered on 13 june 2023.An in-vitro quality control test on sensor (b)(6) did not find any performance issue.However, measurements from the sensor manager software showed led flags were triggered at field currents greater than the established threshold of 412 adc units, which is therefore the root cause of the error.No further investigation was found necessary for this complaint.As part of resolution, the customer was given a sensor replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown 20876-7005
MDR Report Key19009295
MDR Text Key338992198
Report Number3009862700-2024-00461
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491023308
UDI-Public817491023308
Combination Product (y/n)Y
Reporter Country CodeSP
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/12/2023
Device Model Number101967-950
Device Catalogue NumberFG-4500-50-302
Device Lot NumberWP09042
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-