Catalog Number MCM30
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, two clips were pushed as one fire, and the clips were deformed.No patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 4/9/2024.D4: batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: "did more than one clip feeds into the jaws? did device not deploy clip from jaws upon firing? did device fire scissored clips? this may also include ribbon shaped or x-shaped.Did device fire malformed clips?pear/tear drop shape or clip gap?" this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 5/9/2024.D4: batch # a9cj52.Investigation summary : the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcm30 device was returned with a clip in jaws and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining sixteen(16) clips as intended.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key19069965
MDR Text Key339793614
Report Number3005075853-2024-02704
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036002482
UDI-Public10705036002482
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCM30
Device Lot Number386C56
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2024
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-