• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87035
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported that during a pulmonary vein isolation (pvi) procedure to treat an atrial fibrillation (afib), an intellamap orion high resolution mapping catheter and a farawave catheter were selected for use.The orion catheter was used for premap and worked perfectly.Ablation was done with the farawave catheter and then the orion catheter was inserted back into the sheath and deployed into basket mode.A map was made of the right pulmonary veins and then the physician noticed the slider for deploying and undeploying the orion catheter no longer worked.The physician felt unsafe, so catheter was replaced, and they were able to finish mapping.The procedure was able to be completed successfully without any patient complications.The orion catheter is expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key19080659
MDR Text Key339910996
Report Number2124215-2024-21347
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number87035
Device Catalogue Number87035
Device Lot Number0031048166
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-