It was reported that following an elective transcatheter aortic valve replacement procedure, the patient had ventricular fibrillation and ventricular tachycardia and was placed on extracorporeal membrane oxygenation.After the patient was initiated on the motor, there was an m4 motor alarm.The motor operated as intended, except when the motor cable was manipulated.When the motor cable was manipulated, the blood pump sounded ¿off balance¿ and made a noise, and then the patient flows would decrease a little.As long as the motor cable was not manipulated, the motor worked.The motor was changed out very soon after this issue was noticed; the patient did not spend a lot of time using the affected motor.The patient passed away a few hours after the motor issue was noted.The death was not considered to be related to centrimag motor failure.
|