Catalog Number LX110
Device Problems Unintended Ejection (1234); Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 4/11/2024 b3: unknown, assumed first day of month that complaint was reported d4: batch # unk additional information was requested and the following was obtained: "how did device not work? some fell out , some scissored did device jammed (not fire clips)? jamming wasn¿t a problem, clips fell out did device not feed clips? manual did device drop or eject clips? yes did device sideways feed clips? no did device fire malformed clips? yes did device fire scissored clips? yes" an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
 
Event Description
It was reported that during an unknown procedure, the clips are not clipping properly and are not effective.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key19091846
MDR Text Key340689381
Report Number3005075853-2024-02837
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036012863
UDI-Public10705036012863
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLX110
Was Device Available for Evaluation? No
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage Initial
Patient Sequence Number1
-
-