• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; Pulse generator, permanent, implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; Pulse generator, permanent, implantable Back to Search Results
Model Number PM2240
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 04/10/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned without specific complaint or event.As received, the device had normal telemetry communication and no output.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.Device image showed elevated leakage current.The device was cut open to enable further testing and battery was found depleted.Hybrid circuitry was tested by connecting to an external power source.Test results indicated normal current drain.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on (b)(6) 2021.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19102082
MDR Text Key340805284
Report Number2017865-2024-38321
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public(01)05414734507073(10)A000036144(17)180831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2018
Device Model NumberPM2240
Device Lot NumberA000036144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
-
-