During pre-decontamination evaluation of the returned devices, it was observed that two valve struts of the first 29mm sapien 3 ultra resilia were bent outward on the inflow side.As reported by an edwards lifesciences field clinical specialist, patient underwent left carotid approach via surgical cutdown for a transcatheter aortic valve replacement procedure with a 29mm sapien 3 ultra resilia.Utilizing a 16fr esheath+ and 29mm commander delivery system, the valve was noted to buckle/dive in the sheath shaft, requiring removal of the system.The valve was exiting the out of the side of the sheath via a puncture in the liner.The devices were removed as a unit.A new system was prepped, and a new 29mm sapien 3 ultra resilia was successfully delivered.At the end of the procedure upon access site closure a left carotid artery dissection was noted.The dissection required graft and stenting.Procedure was performed successfully, and the patient was transferred in stable condition out of the or.During pre-decontamination evaluation of the returned devices, it was observed that two valve struts were bent outward on the inflow side.
|