• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; TEST, HIV DETECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; TEST, HIV DETECTION Back to Search Results
Catalog Number 7D2648
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
The customer reported twenty-three (23) false positive results with the determine hiv 1/2 ag/ab tests performed on unknown dates with unknown sample types.This manufacturer's report addresses test twelve (12) of twenty-three (23) and is associated with lot# 802592 (qty 4).The customer indicated that confirmation pcr testing was performed on 1 of the 23 tests on an unknown date and generated a negative result.It is unknown if confirmation or repeat testing was performed for the other 22 results.Although requested, no additional patient information, including treatment and outcome, was provided.
 
Manufacturer Narrative
B3: the date indicated is an approximation as the exact event date was not provided.The remainder of the investigation remains in progress.A supplemental report will be provided after completion or upon receipt of new information.
 
Manufacturer Narrative
B3: the date indicated is an approximation as the exact event date was not provided.Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 802592 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 7d2648 / lot 802592 and test base part number 10732998 / lot 794332.The lot met the required release specifications.A review of the complaints reported false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 802592 showed that the complaint rate is (b)(4).In conclusion, abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue.
 
Event Description
The customer reported twenty-three (23) false positive results with the determine hiv 1/2 ag/ab tests performed on unknown dates with unknown sample types.This manufacturer's report addresses test twelve (12) of twenty-three (23) and is associated with lot# 802592 (qty (b)(4)).The customer indicated that confirmation pcr testing was performed on 1 of the 23 tests on an unknown date and generated a negative result.It is unknown if confirmation or repeat testing was performed for the other 22 results.Although requested, no additional patient information, including treatment and outcome, was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T
Type of Device
TEST, HIV DETECTION
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key19149296
MDR Text Key340905207
Report Number1221359-2024-00351
Device Sequence Number1
Product Code MZF
UDI-Device Identifier00811877011101
UDI-Public010081187701110110802592
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BP120037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7D2648
Device Lot Number802592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-