Catalog Number UNK ADAPTOR |
Device Problems
Device-Device Incompatibility (2919); Naturally Worn (2988)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
There have been some issues with the mi reamers.There seems to be 2 separate issues, one where the 'elbow' joint in the reamer itself is very stiff and therefore not allowing the grater to spin freely and a seperate issue where the connection between the reamer and the adaptor is getting stuck and difficult to remove and when removed there is clear signs of wear on the reamer and inside the adapter.We have been advised by the power tool team that these components are all compatible, but this is becoming a serious problem for the hospital as they are not able to use them and are having to open multiple sets per case which is causing delays during surgery.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown, therefore a¿device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|
|
Search Alerts/Recalls
|
|