• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO NC SPRINTER RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO NC SPRINTER RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, Back to Search Results
Model Number NCSP2521X
Device Problems Burst Container or Vessel (1074); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
During a procedure an attempt was made to use one nc sprinter coronary balloon catheter.The device was inspected with no issues.Negative prep was performed with no issues.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that a balloon burst, or leak occurred during balloon inflation.The patient is alive with no injury.
 
Manufacturer Narrative
Product analysis: the device was received for analysis.The device returned with the balloon in a post inflated state.The balloon failed negative prep.On pressurisation of the device liquid was observed exiting a tear at the exchange joint.The material at the tear site was jagged and uneven.No other damage evident to the remainder of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC SPRINTER RX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19180508
MDR Text Key341020818
Report Number9612164-2024-01991
Device Sequence Number1
Product Code LOX
UDI-Device Identifier00643169282322
UDI-Public00643169282322
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P790017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNCSP2521X
Device Catalogue NumberNCSP2521X
Device Lot Number227806047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/01/2024
Date Device Manufactured11/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-