• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-XTW
Device Problems Break (1069); Difficult to Remove (1528); Difficult to Open or Close (2921)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 04/02/2024
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2024, a mitraclip procedure was performed to treat mixed mitral regurgitation (mr) grade 4.A xtw (lot: 31026a2062) was chosen for implantation.The clip was advanced to the valve.During initial grasping attempts, the clip became stuck in the chordae.The clip was able to be freed via troubleshooting but chordal damage had occurred.The patient's pre-exisitng flail increased.Once free the clip grasped the leaflets again, but one of the grippers was not functioning when attempting to raise.The clip was removed from the patient where the gripper line was observed to be broken.No replacement device was used and the procedure was aborted with mr unchanged grade 4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19182509
MDR Text Key341054235
Report Number2135147-2024-01845
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public(01)05415067037381(17)241025(10)31026A2062
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number31026A2062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Other;
Patient Age68 YR
Patient SexMale
Patient Weight111 KG
-
-