Model Number CD2357-40C
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The device was received without a reported complaint.A failure event was observed during analysis.Final analysis found the battery depletion was premature based on the device usage.Device image indicated normal current drain during implant period and no high current drain sources were found.The battery analysis by the manufacturer found an internal battery anomaly to be the cause of premature battery depletion.No other anomalies were detected.
 
Event Description
This report is to advise of a failure event that was observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19194201
MDR Text Key341434100
Report Number2017865-2024-39862
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508063
UDI-Public(01)05414734508063(10)A000035352(17)190228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/28/2019
Device Model NumberCD2357-40C
Device Lot NumberA000035352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2023
Was the Report Sent to FDA? No
Date Manufacturer Received04/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-