Catalog Number 1601-08127
Device Problem Corroded (1131)
Patient Problem Discomfort (2330)
Event Date 11/15/2023
Event Type  Injury  
Event Description
It was reported by claimant's counsel that the patient underwent left tha on (b)(6) 2014, and was implanted with an lfit v40 femoral head and a secur fit advanced stem.Recently he felt something moving and presented with left hip discomfort.He was revised on (b)(6) 2023, and intraoperatively the surgeon noted "trunnion was visualized and was corroded and deformed".
 
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
127 SIZE 8 SECUR-FIT ADVANCED STEM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key19215331
MDR Text Key341419600
Report Number0002249697-2024-00670
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327021318
UDI-Public07613327021318
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Catalogue Number1601-08127
Device Lot NumberMMED8W
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/05/2024
Initial Date FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/18/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexMale
-
-