• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 6 SAGITTAL SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 6 SAGITTAL SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 6208000000
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 07/11/2013
Event Type  malfunction  
Event Description
It was reported that upon receipt at the user facility the device was leaking an unknown substance resembling blood.As this was an out-of-box failure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
The reported leak was confirmed by a manufacturer repair technician.Upon evaluation, it was confirmed that the leaking substance was excess lubricant, which can be caused by a mechanical issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 6 SAGITTAL SAW
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kate longman
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3551234
MDR Text Key19732460
Report Number0001811755-2014-00022
Device Sequence Number1
Product Code KIJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/11/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6208000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2013
Date Manufacturer Received07/11/2013
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-