STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 6 SAGITTAL SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 6208000000 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/11/2013 |
Event Type
malfunction
|
Event Description
|
It was reported that upon receipt at the user facility the device was leaking an unknown substance resembling blood.As this was an out-of-box failure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The reported leak was confirmed by a manufacturer repair technician.Upon evaluation, it was confirmed that the leaking substance was excess lubricant, which can be caused by a mechanical issue.
|
|