• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG COROX OTW-L 85-BP; LV LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG COROX OTW-L 85-BP; LV LEAD Back to Search Results
Model Number 368346
Device Problem Programming Issue (3014)
Patient Problem Muscle Stimulation (1412)
Event Date 12/17/2013
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.The analysis is therefore based on the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process which might be related to the clinical observation.In summary, the device was not returned for analysis.The review of the quality documents confirmed a regular device manufacturing.
 
Event Description
This patient was experiencing diaphragmatic stimulation from this lead.The output was lowered which resolved the issue.This lead remains implanted and no further adverse events have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COROX OTW-L 85-BP
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-12 359
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key3564657
MDR Text Key11541604
Report Number1028232-2013-03988
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/30/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number368346
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age68 YR
-
-