• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM Back to Search Results
Catalog Number 1011540-18
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 12/18/2013
Event Type  Injury  
Manufacturer Narrative
(b)(4).A review of the lot history record revealed no non-conformances that would have contributed to the reported event.Dissection is listed in the herculink elite stent system instructions for use as a known potential adverse event.Based on the information reviewed, there is no indication of a product deficiency.
 
Event Description
It was reported that a herculink elite 7x18 stent was deployed in right renal artery to revascularize an 80% stenosis.Reportedly, after stent deployment, a dissection was detected in the distal part of the vessel.Another herculink elite 6x18 was deployed to cover the dissected area.There were no reported adverse patient sequelae and no reported significant clinical delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device is expected to be returned for evaluation.It has not yet been received.A follow-up will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Type of Device
STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key3566491
MDR Text Key11635460
Report Number2024168-2014-00172
Device Sequence Number1
Product Code NIN
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/18/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Catalogue Number1011540-18
Device Lot Number2071961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-