Model Number IMMULITE 2000 ANDROSTENEDIONE |
Device Problem
High Test Results (2457)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/09/2013 |
Event Type
malfunction
|
Event Description
|
The customer has observed high bias on neqas survey samples for the androstenedione assay on an immulite 2000 instrument, when using reagent lot 315.The customer has observed that the higher the value for androstenedione, the higher the bias observed.The five survey samples from four different specimen pools were tested four times, in (b)(6) 2012, (b)(6) 2013.There were no known reports of patient intervention or adverse health consequences due to the high bias observed on the neqas survey samples when using reagent lot 315.
|
|
Manufacturer Narrative
|
Siemens has investigated the incidence of over recovery of patient samples and quality controls on the immulite/immulite1000 androstenedione assay lot number 409 and above, and immulite 2000/immulite 2000 xpi androstenedione assay lot number 314 and above.Siemens has confirmed that these lots are linear up to a concentration of 5.5 ng/ml, instead of the 10 ng/ml as stated in the instructions for use (ifu).Quality control materials at concentrations between 5.5 ng/ml and 10 ng/ml will detect this issue.An urgent medical device correction (umdc) 2013-12-19 was sent to us customers and an urgent field safety notice (ufsn) 4008 was sent to ous customers in december of 2013.The umdc and ufsn state that customers are requested to use quality control materials with at least two levels having concentrations less than 5.5 ng/ml, and that patient values >5.5 ng/ml should be verified using an alternate method.
|
|
Manufacturer Narrative
|
The original mdr 2432235-2014-00062 was filed on january 10, 2014.Additional information (5/27/14): the over recovery and loss of linearity of the immulite systems androstenedione assay has been identified to be variability of a critical raw material (variation in the activity of rma160 alkaline phosphatase (unlabeled/low activity alkaline phosphatase, cold ap) in the reagent wedge).In addition, siemens healthcare sent out a customer bulletin 2014-05 in may 2014, announcing the restoration of the linearity of the androstenedione assay on the immulite 2000/immulite 2000 xpi to 10 ng/ml (35 nml/l), beginning with kit lot 331.
|
|
Search Alerts/Recalls
|