Within the first 30 minutes of treatment, the nurse visualized foam with micro bubbles in the venous line when she went to administer the patient¿s medication.The patient complained of feeling anxious and short breath, treatment was stopped, and the blood from the extracorporeal circuit was not returned to the patient.The nurse does not recall if the phoenix generated the air in blood alarm or not.The patient was transferred to the hospital and diagnosed with ¿an air embolus which absorbed" the patient was later discharged with no sequelae.Limited clinical information related to the patient has been provided.The customer recently switched from gambro dialyzers to a competitor¿s dialyzer.The priming technique differs between the two dialyzers.Gambro¿s request for an onsite visit by gambro's clinical educator was denied by the director of the dialysis unit.The blood tubing set was discarded and not available for investigation.The phoenix machine was inspected by a gambro technical service representative and within specification.The air in blood sensor was proactively replaced and returned to the manufacturer for analysis.The air in blood sensor was inspected and determined to be operating within specification and as intended.
|