Result: the catheter appears to be fractured approximately 33.0 cm from the hub in proximal shaft.The break is in the middle of a material extrusion and not at a material bond.Conclusion: the device was returned for analysis and has been evaluated.The complaint indicates that the catheter was defective when it was remove from the box and noticed during preparation.Evaluation of the returned product confirmed that the catheter was fractured in the proximal shaft.Based on the description of the complaint, the cause of this damage could not be determined however, if the device is mishandled during removal from the packaging or preparation for use, damage such as this may occur.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|