Catalog Number 5100088000 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/18/2013 |
Event Type
malfunction
|
Event Description
|
The tps micro driver was returned to stryker instruments for service, and during functional evaluation it was noted that the device ran in safe mode.There was no patient involvement and no adverse consequences associated with the device.
|
|
Event Description
|
The tps micro driver was returned to stryker instruments for service, and during functional evaluation it was noted that the device ran in safe mode.There was no patient involvement and no adverse consequences associated with the device.
|
|
Manufacturer Narrative
|
Failure analysis is ongoing; additional information may be submitted once the results analysis is complete.
|
|
Manufacturer Narrative
|
The reported event of running in safe mode was duplicated, and the technician noted a corroded pcb which can cause this failure.
|
|