• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. PARADYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. PARADYM Back to Search Results
Model Number PARADYM RF DR 9550
Device Problems Under-Sensing (1661); Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2013
Event Type  Injury  
Event Description
Reportedly, on (b)(6) 2013, the icd and the atrial lead were replaced.Few days later, the remote monitoring report showed atrial lead impedance below 200 ohms, and sensing decrease.However, no alert was raised.On (b)(6) 2013, the device was interrogated with the programmer and atrial impedance was found at 283 ohms.Preliminary analysis suggested a lead or connection issue at atrial side.
 
Manufacturer Narrative
This event concerns a device that was manufactured and used outside the united states.Analysis is pending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADYM
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L.
saluggia
IT 
Manufacturer (Section G)
SORIN GROUP ITALIA S.R.L.
via crescentino s.n.
saluggia (vc) 1304 0
IT   13040
Manufacturer Contact
e vincent
4 avenue reaumur
clamart cedex 92140
FR   92140
46013687
MDR Report Key3580830
MDR Text Key4138361
Report Number1000165971-2014-00004
Device Sequence Number1
Product Code MRM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2014,12/13/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/01/2014
Device Model NumberPARADYM RF DR 9550
Device Catalogue NumberPARADYM RF DR 9550
Device Lot Number2683
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received12/13/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-