Valve implanted in (b)(6) 2013, now clinically overdrainage, valve pressure position at 200 mm h20, intraop testing of the valve showed a pressure of 80 mm h2o explantation.
|
The suspected valve has been received and analyzed by our quality control laboratory.According to the device inspection results, before and after cleaning, the valve did not meet all of its functional specifications and the difficulty to undergo the highest pressure setting procedure has been duplicated.Indeed, 2/3 of the measured values were outliers.However, no abnormal valve behaviour was observed when the pressure setting procedure was operated in laboratory environment to adjust the valve on the lower pressure positions and back to the higher pressure position.Regarding the closing pressure on high pressure setting, the valve meets its requirements specifications with a conforming hysteresis from the operating pressure.The main explanations for this discalibration might be the post hemorrhagic csf which could have modified the ball-in-cone mechanism and the surface of the rub ball and its metallic seat.
|