Review of the device manufacturing record history confirmed device met pre-release specifications.Engineering evaluation state the gore viabahn endoprosthesis is compressed distally approximately 5.5cm.Some expansion was noted at the proximal end of the endoprosthesis.The deployment line is severed and approximately 1.5cm of the compressed outer layer of zipper is deployed.There is also a kink in the dual lumen approximately 38cm from the hub.Based on the examination performed, no anomalies were found that could be attributed to the mfr of the device.The warnings section of the ifu state: do not withdraw the gore viabahn endoprosthesis with heparin bioactive surface back into the introducer sheath once the endoprosthesis is fully introduced.Withdrawing the gore viabahn endoprosthesis with heparin bioactive surface back into the sheath can cause damage to the endoprosthesis, premature deployment, deployment failure, and / or catheter separation.If removal prior to deployment is necessary, withdraw the gore viabahn endoprosthesis with heparin bioactive surface to a position close to but not into the introducer sheath.Both the gore viabahn endoprosthesis with heparin bioactive surface and introducer sheath can then be removed in tandem.After removal, do not use the gore viabahn endoprosthesis with heparin bioactive surface oar introducer sheath.
|