• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSOLUTE PRO .035 SELF EXPANDING STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSOLUTE PRO .035 SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number 1011929-100
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Patient Involvement (2645)
Event Date 12/26/2013
Event Type  malfunction  
Event Description
It was reported that while opening the device package, the inner pouch was not sealed.Another device was used completing the procedure.There was no patient involvement and no reported clinically significant procedural delay.There was no additional information provided.
 
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.Although the pouch was returned with the seal opened, there was evidence on the pouch where it had been previously sealed.Based on a visual analysis of the returned product, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.The results of the query of similar incidents in the complaint handling database for this lot did not indicate a manufacturing issue.Based on the reviewed information, no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO .035 SELF EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key3583909
MDR Text Key4192755
Report Number2024168-2014-00364
Device Sequence Number1
Product Code FGE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K072708
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/26/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Catalogue Number1011929-100
Device Lot Number2062661
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/09/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/04/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-