• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY ULTRAFLEX? ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY ULTRAFLEX? ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00513730
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/23/2013
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultraflex esophageal covered stent was implanted during a stent placement procedure performed on (b)(6) 2013.According to the complainant, the stent was being implanted due to a 15 cm malignant stricture in the esophagus.Reportedly, the patient¿s anatomy was not tortuous and had not been dilated prior to stent placement.No issues were noted to the device prior to the procedure.During the procedure, the stent was deployed under fluoroscopy.The physician inserted an endoscope to check the position of the stent and noted that the cover was missing from the stent.The stent remains implanted to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Reported event of stent cover missing.The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX? ESOPHAGEAL NG
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
ingrid matte
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key3584789
MDR Text Key4035792
Report Number3005099803-2014-00272
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K091816
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/19/2014
Device Model NumberM00513730
Device Catalogue Number1373
Device Lot Number0014972420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-