Catalog Number 413.020
Device Problems Break (1069); Detachment Of Device Component (1104); Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2013
Event Type  malfunction  
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: during a removal osteosynthesis surgery, it was impossible to unscrew the screws of the plate (screws blocked).The screw heads were destroyed with a drill in order to remove the plate.There was an extension of delay of surgery.Fragments remain in the soft tissue.No clinical consequence in intraoperative.This is report 3 of 4 for complaint (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Date of implant unknown.Investigation could not be completed and no conclusion could be drawn as no device was returned and no lot number was provided.Placeholder.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
An evaluation was performed on the returned product.The received screw was drilled out, head is not present.The relevant dimensions of the locking screws can not be measured as the screw heads were drilled out for removal.Also the locking threads in the plate can not be evaluated due to damage during removal.The examination of the raw-material testing certificates and the manufacturing records showed no deviations regarding material analysis, strength and structural stability at the time of manufacture.The damage can be contributed to excessive force that was required for removal.Therefore, this complaint has been deemed indeterminate from a manufacturing perspective.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKSCR Ø3.5 SELF-TAP L20 TAN
Manufacturer (Section D)
SYNTHES USA
west chester PA
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
8006207025
MDR Report Key3595228
MDR Text Key20106602
Report Number2520274-2014-00330
Device Sequence Number1
Product Code KTT
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K000684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number413.020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/18/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Patient Sequence Number1
-
-