AV-TEMECULA-CT XACT CAROTID STENT SYSTEM Back to Search Results
Catalog Number 82097-01
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/15/2014
Event Type  malfunction  
Event Description
It was reported that when the 10 x 30 x 136 xact self expanding stent system (sess) was un-packaged, it was found that 2mm of the stent was exposed.The handle was rotated back, but the stent could not be re-sheathed.The decision was made to use a new sess.The technician then deployed the stent on the back table.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the complaint handling database for the reported lot revealed no other incidents.Based on the information reviewed, there is no indication of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Type of Device
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key3608927
MDR Text Key4088443
Report Number2024168-2014-00629
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2016
Device Catalogue Number82097-01
Device Lot Number3042261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-