Catalog Number 234-200-200 |
Device Problem
Malposition of Device (2616)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/14/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that a doctor was using the microfx drill in the hip with a 90 degree curve guide.In his first attempt to drill a hole, the drill snapped where the bendable portion attaches to the rigid shaft.The doctor drilled at a controlled speed and had complete control of the guide with zero movement.
|
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that a doctor was using the microfx drill in the hip with a 90 degree curve guide.In his first attempt to drill a hole the drill snapped where the bendable portion attaches to the rigid shaft.The doctor drilled at a controlled speed and had complete control of the guide with zero movement.
|
|
Manufacturer Narrative
|
The product was returned for investigation.The reported failure mode was confirmed.The drill is broken at the proximal end of the cable where the cable meets the rigid shaft of the drill.The cable appears to be unraveling near the drill tip and the drill tip shows wear along the cutting edges.Based on the event description and additional information provided, hard bone as well as user technique leading to skiving are considered the primary root causes.In sum, the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
|