Model Number T20A2U |
Device Problem
Reset Problem (3019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/11/2013 |
Event Type
Injury
|
Event Description
|
It was reported that the device experienced a power on reset (por) that required a manual guided reset (mgr).It was noted that the por was most likely due to radiation.The mgr was successful and the device remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
This event occurred outside the us.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.This model number is not approved for distribution in the united states, however, it is similar to a device marketed in the u.S.The event is being reported due to an alleged malfunction.(b)(4).
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a partial electrical reset.The device partial reset was due to a watchdog timeout (wdto status) that occurred 2010-(b)(6), likely bit flip.Device partial reset, but device started logging data into incorrect memory addresses due to an ecc log pointer error.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|