We received one single dpt kit.Examination revealed that a section of the dpt cable cover appeared to be burned and melted.The burn section was approximately 1" in length and 1" proximal from the dpt housing.It was also noted that the number 5 leadwire was exposed at the burned area.The cable cover turned black at the center of the burn and brown in the area surrounding the burned mark.Further examination revealed that the cable turned to a 90 degree angle at the burned area.However, x-rays showed that the leadwires were intact.The leadwire maintained separated from one another and no short condition between the leadwires at the burn area was observed.The x-rays also showed that the rest of the dpt circuit, which includes the dpt cable connector, cable and the sensor chip were intact and there was no short condition.It was also revealed that the dpt zeroed and sensed pressure accurately on the pressure monitor.The electrical testing also showed that the dpt electronic components were intact because both input impedance and output impedances were within specifications.Additionally, as stated on the edward's website, these devices and their associated cables are not intended for use inside the bore of the mr system and should not be in contact with the patient during the mr procedure.
|