Device evaluated by manufacturer: device was retuned for analysis.A visual and microscopic examination was performed and found that the radiopaque distal tip of the protection wire was found slightly curved.It was not wavy, or stretched.The filter bag was found retracted into the delivery sheath with 4 mm of the nosecone exposed.The filter bag was successfully deployed and then retracted; the filter bag was observed to be in good condition and met specification.The peel away test was also performed successfully, no anomalies were observed.The inner wire was found broken at the distal solder joint, but not detached.Three mm of the wire was found protruding out of the coil.No discoloration was observed indicating overheating during soldering.Evidence of solder on the broken end of the wire was found, and no kinks or dents that could have weakened the wire were found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|