• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Device Operates Differently Than Expected (2913); Battery Problem: High Impedance (2947)
Patient Problem Inadequate Pain Relief (2388)
Event Date 01/07/2014
Event Type  Injury  
Event Description
It was reported the pt is not feeling stimulation.The sjm rep met with the patient and diagnostic testing revealed high impedance readings.Reprogramming was unable to resolve the issue.X-rays were taken and no anomalies were identified.Surgical intervention may be taken at a later date as the next course of action.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
toi milner
6901 preston rd.
plano, TX 75024
9723098107
MDR Report Key3634715
MDR Text Key19199907
Report Number1627487-2014-15073
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 01/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/01/2006
Device Model Number3186
Device Lot Number27083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS IPG: MODEL 3788
Patient Outcome(s) Other;
Patient Age49 YR
-
-