Catalog Number 1074350-18 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/24/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that upon removal of a 3.5x18mm xience xpedition stent delivery system (sds) from the protective hoop, a small piece of plastic came off the stent.Reportedly, this plastic is thought to have been the tip of the sds but this was not confirmed.The device was set aside for return and not used.There was no patient involvement and no clinically significant procedural delay.There was no additional information provided.
|
|
Manufacturer Narrative
|
(b)(4).Evaluation summary: the device was returned for analysis.The foreign material was unable to be confirmed.Based on a visual inspection of the returned device, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the complaint handling database for the reported lot revealed no other incidents reported for foreign material.Based on the reviewed information, no product deficiency was identified.
|
|