Device was used for treatment, not diagnosis.510k#: device is not distributed in the united states, but is similar to device marketed in the usa a review of the device history records was performed and no complaint related issues were found.The visual inspection of the returned device by the complaint handling unit revealed all four wings of the peek implant are still un-deployed.Additionally circular marks are clearly visible on the connecting end of the implant.The review of the manufacturing documents revealed that the implant was manufactured in april 2013 according to the specifications.No manufacturing related conditions were found.The complaint condition is likely the result of the implant not attached correctly onto the implant holder (the slots of the implant did not match with the tappets of the implant holder), resulting in a wrong orientation of the implant and finally preventing the deployment of the wings as the wings were covered by the wall of the sleeve.No product fault could be detected.Placeholder.
|