Upon receipt at medtronic's quality laboratory, kinks and telescoping were observed along the delivery catheter system (dcs) capsule and kinks were observed at the proximal end of the capsule.The handle was intact.The micro knob (thumb wheel) retracted and advanced the capsule.The macro slider (cursor) moved to fully advanced and retracted positions and locked in place when released.The distal tip of the capsule seated flush against the plunger tip when fully advanced.Deformation (gouge marks) on both catheter tabs appeared to be due to wear from the valve frame loop upon deployment.The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The observed telescoping and the kinks on the capsule are consistent historically with devices tracked over an aortic arch and are not indicative of a device issue.The deformation of the dcs tabs showed evidence of a tensile force applied, most likely from the frame loops of the valve.The analysis results confirm the clinical observation that the user attempted to retrieve a partially-deployed valve (one that is not fully captured in the capsule) that resisted retrieval, and the user applied tension in pulling back the dcs in the retrieval attempt, which caused damage to the dcs tabs.This damage to the dcs tabs, along with continued tension on the system, allowed the valve frame loops to release from the tabs.The dcs is not designed to be a retrieval device; the instructions for use (ifu) states: "once deployment is initiated, retrieval of the bioprosthesis from the patient (e.G., use of the catheter) is not recommended.".
|