• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500 Back to Search Results
Catalog Number 10492730
Device Problem Maintenance Does Not Comply To Manufacturers Recommendations (2974)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/02/2013
Event Type  Other  
Manufacturer Narrative
Operator was running patient samples event when the automatic quality control (aqc) was off.They should not have run patient samples during when aqc was off.Operator is being instructed to enable the aqc and advised to secure that screen, so no one can go in and disable it again.The event has occurred due to an operator error.
 
Event Description
Customer reported that aqc (automatic quality control) was not run on the instrument from (b)(6) 2013.Customer indicated that 31 patient samples were run during this time frame.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD
northern rd
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steve andberg
2 edgewater dr.
norwood, MA 02062
7812693655
MDR Report Key3651383
MDR Text Key19389219
Report Number1217157-2014-00023
Device Sequence Number1
Product Code KHP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/08/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-