Catalog Number 6605632
Device Problems Detachment Of Device Component (1104); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2014
Event Type  malfunction  
Event Description
A customer reported that, after running controls at startup, they received retic control vote outs on the lh750 instrument.A field service engineer (fse) was dispatched.When the fse arrived on site, he noticed a leak of blue violet fluid from disconnected tubing that had popped off a check valve (cv103b).The volume was reported as 20 ml and the leak was not contained.No erroneous patient results were generated in connection with this event.
 
Manufacturer Narrative
The field service engineer (fse) replaced the tubing that had popped off at check valve 103 b and the check valve to resolve the leak.Sample lines and a t fitting were also replaced to resolve the low retic values recovered on the controls.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER® LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key3673600
MDR Text Key4345695
Report Number1061932-2014-00545
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-