Refer to evaluation summary manufacturer ref # (b)(4).It was reported that during the second transeptal, the left atrium was punctured and the patient's blood pressure dropped.Ice confirmed a pericardial effusion.A pericardiocentesis was performed extracting 700 cc's of fluid and the patient stabilized.The returned device was visually inspected upon receipt and the inner section of the luber hub had some dark ¿ red material that appeared to be human blood.Per the event, the catheter was tested for electrical performance, temperature response and stockert compatibility and it was found within specifications.Furthermore, a deflection test was performed and the catheter passed.The catheter was also evaluated for eeprom, carto 3, 4 khz and calibration functionality.The catheter was recognized by carto 3 system, no error messages were displayed and the catheter was properly visualized.Eeprom data demonstrates the catheter was properly calibrated during manufacturing.Finally, an irrigation test was performed and the catheter failed since three out of six irrigation ports were found occluded with dark ¿ red material, which appeared to be human blood.However, it remains unknown the origin of that material.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.The catheter failed the irrigation test, but the root cause of the pericardial effusion remains unknown.The ifu states that careful catheter manipulation must be performed in order to avoid cardiac damage, perforation or tamponade.
|
On march 14, 2014, upon receiving the product in biosense webster lab, it was noticed that inside luber hub connector irrigation tubing appears occluded.Investigation is still in progress.A supplemental report on device evaluation will be submitted.(b)(4).
|