Catalog Number S-1510TD
Device Problems Break (1069); Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2014
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure the "tip" of the cutter device "broke off".The report indicated that "when the surgeon tried to make five or more holes in bone flap for tenting, the tip of the twist drill in question broke off".The report indicated that the "broken tip of twist drill was removed without any problem".It was further clarified that since tenting was performed out of operating field, the piece of the device did not fall into the patient's skull.There were no reports of injuries, medical intervention or prolonged hospitalization.There was no delay to the planned surgical procedure.It was unknown if a spare device was available for use.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
(b)(6).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.  reliability engineering evaluated the device.  a visual inspection of this device was performed under 10x magnification which confirmed the reported condition.  the assignable root cause was determined to be exposure to excessive lateral force during use. the device met all dimensional specifications.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.55MM X 10MM FLUTED TWIST DRL
Type of Device
MOTOR, DRILL, ELECTRIC - CUTTING BURR
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key3677062
MDR Text Key4302086
Report Number1045834-2014-11646
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberS-1510TD
Device Lot NumberG273080945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
-
-