U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES MONUMENT 24MM COCR RADIAL HEAD; PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MONUMENT 24MM COCR RADIAL HEAD; PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER Back to Search Results
Catalog Number 09.402.024S
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 02/19/2014
Event Type  malfunction  
Event Description
During radial head prosthesis the set screw broke off while the surgeon was attaching the radial head to the radial stem.The screwdriver is supposed to have a torque limiter attached while inserting the set screw.The limiting device was not attached.The screw head snapped at the head and was removed from the screwdriver tip, and is now available for evaluation.The surgeon continued with the procedure with only the screw shaft securing the head and stem.The procedure was completed to the satisfaction of the surgeon.The screw shaft remains embedded in the radial head, part number 09.402.024s, joining the radial head and radial stem.The surgeon completed the procedure successfully without delay and without any adverse event to the patient.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
A review of the device history records has been requested.The investigation could not be completed and no conclusion could be drawn as no device was returned.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional information provided.Nemcomed (presently avalign technologies ¿ nemcomed) manufactured the 24mm cocr radial head standard height, p/n 04.402.024s, lot number 6955108, (b)(4).There were no mrrs, ncrs, or complaint-related issues during the manufacture of the product that would contribute to this complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
24MM COCR RADIAL HEAD
Type of Device
PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER
Manufacturer (Section D)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer (Section G)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3677715
MDR Text Key4258759
Report Number1719045-2014-10053
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK112030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number09.402.024S
Device Lot Number6955108
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/17/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2012
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age37 YR
Patient Weight100
-
-