U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR CP
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/14/2014
Event Type  malfunction  
Event Description
Discordant, falsely low alpha-fetoprotein (afp) results were obtained on one patient sample upon a 1:20 dilution and neat repeat on an advia centaur cp instrument.The discordant results were not released to the physician(s), as they did not match the patient¿s clinical history.The sample was repeated on the same instrument on a 1:10 dilution and post service, resulting as expected.The sample was also sent for testing at an alternate lab and the result matched the patient¿s clinical history.It is unknown which result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low afp results.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse performed a total service visit.The cse replaced the sample syringe and performed a database cleanup, deleting results greater than 30 days old.Quality controls and a patient sample were run, resulting within range.The cause of the discordant, falsely low afp results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR CP
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
STRATEC BIOMEDICAL SWITZERLAND AG
registration #: 3008494306
registration number: 8020888
gewerbestrasse 6, neuhausen am rheinfall 8212
SZ   8212
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key3678795
MDR Text Key4241950
Report Number2432235-2014-00231
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR CP
Device Catalogue Number086-A002-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse