A steris service technician arrived at the facility, inspected the v-pro 1 plus sterilizer and confirmed the equipment was operating to specification.The reported event could not be duplicated and no repairs to the v-pro 1 plus sterilizer were required.The employee subject of the reported event was not wearing the appropriate ppe as recommended in the v-pro 1 plus operator manual.The operator manual (6-11) states, "unload sterilization unit: steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical -resistant gloves when using the sterilization unit." in addition, the operator manual (1-2) states, "danger - chemical injury hazard: any visible liquids in the chamber must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions." the steris account manger is scheduled to visit the facility on monday, (b)(4) 2014 in order to provide additional in-service training to the facility staff on proper use of ppe.
|